鐘金德 郭貴勇
【摘 要】本文依據JJG 139-2014拉力、壓力和萬能試驗機檢定規(guī)程,且對各不確定度分量主要來源進行分析與評定,再按材料試驗機測量結果的數據模型對其合成不確定度進行評定,結合具體數據對各不確定度分量進行計算并得出合成不確定度和擴展不確定度。
【關鍵詞】材料試驗機;測量結果;不確定度評定
中圖分類號: TU502.1 文獻標識碼: A 文章編號: 2095-2457(2018)33-0030-002
DOI:10.19694/j.cnki.issn2095-2457.2018.33.012
【Abstract】According to the verification regulation of JJG 139-2014 tension、compression and universal testing machines, this paper analyses and evaluates the main sources of uncertainty components, then evaluates the synthetic uncertainty according to the data model of material testing machine measurement results. Combined with the specific data, the uncertainties of each component are calculated, and the combined uncertainties and extended uncertainties are obtained.
【Key words】Material testing machine; Measurement results; Uncertainty evaluation
0 概述
(1)測量方法:依據JJG 139-2014拉力、壓力和萬能試驗機檢定規(guī)程
(2)環(huán)境條件:溫度:(20±5)℃,相對濕度≤80%,無影響正常工作的振動沖擊,無強電磁干擾,標準器供電正常。
(3)測量標準:0.2級標準測力儀,測量范圍:(20~500)kN。
(4)被測對象:微機控制電子式萬能試驗機,型號規(guī)格:WDW-200E;編號:0426。
(5)測量過程:在規(guī)定條件下,開機預熱30min;預加3次額定負荷;清零,按遞增力進行測量,以標準測力儀為準,試驗機檢定過程中以1mm/min速率加載,離檢定點3kN的時候速率切換至0.02mm/min,到達每個檢定點后,記錄試驗機示值,直到本次試驗的最大力值點,然后卸荷;重復進行三組測量。
(6)評定結果的使用:符合上述條件的材料試驗機,一般可直接使用本不確定度的評定結果,其它可參照使用本評定方法。
1 測量模型
2 標準不確定度分量的評定
(1) 輸入量的標準不確定度的u(F)評定:其不確定度主要來源于材料試驗機的重復性,可采用A類方法進行評定u1(F)。還來源于材料試驗機的分辨率,它估計為均勻分布,可以用B類方法評定u2(F)。依據JJF1033-2016檢定或校準結果的重復性分量取重復性分量與分辨力分量的大者。
3 擴展不確定度的報告與表示
當試驗機在100kN測量點時,其示值誤差測量結果擴展不確定度為U=0.24kN;它是由合成標準不確定度uc=0.116kN與包含因子k=2之乘積取得。
當試驗機在200kN測量點時,其示值誤差測量結果擴展不確定度為U=0.47kN;它是由合成標準不確定度uc=0.232kN與包含因子k=2之乘積取得。
【參考文獻】
[1]拉力、壓力和萬能試驗機檢定規(guī)程:JJG 139-2014[S].
[2]計量標準考核規(guī)范:JJF 1033-2016[S].