【摘要】本文根據(jù)尿素檢測儀的測量原理提出了一種測試尿素檢測儀性能的方法,主要檢測其波長與濃度示值誤差,并對檢測結果進行不確定度評定。該測試方法經過實驗驗證,結果準確可靠。
【關鍵詞】尿素檢測儀;測試;不確定度
【DOI編碼】10.3969/j.issn.1674-4977.2024.06.086
Research on the Method of Testing the Performance of Urea Detector Based on Colorimetric
ZHONG Ni, LUO Yilong, REN Xingru, DU Lan, LI Xiangzhao
(Shenzhen Institute of Metrology and Quality Testing, Shenzhen 518055, China)
Abstract: Based on the measurement principle of the urea detector, this paper proposes a method to test the performance of the urea detector, mainly to detect its wavelength and concentration indication error, and to evaluate the uncertainty of the test results. This test method has been verified by experiments, and the results are accurate and reliable.
Keywords: urea detector; testing; uncertainty
尿素含量的主要檢測方法有比色法、色譜法、光譜法以及流動注射法。其中,比色法具有檢出限較低、準確度較低、精密度較高、成本較低、操作過程簡單等特點[1],在泳池水質檢測中得到了廣泛的應用。
尿素檢測儀的性能測試項目主要包括波長示值誤差、儀器濃度示值誤差和儀器重復性。波長示值誤差使用光譜分析儀進行掃描檢測,儀器濃度示值誤差和重復性測試通過水中尿素標準溶液進行檢測。通過對尿素檢測儀的性能進行測試可以讓制造廠家更加了解自身產品的性能,有助于廠家改進產品,提升產品市場競爭力。
1.1測試環(huán)境條件
環(huán)境溫度:15~35℃;相對濕度:≤80%;儀器周圍無環(huán)境干擾和強光干擾。
1.2主要標準器與標準物質
標準物質:水中尿素標準物質,質量濃度為1000 mg/L,相對不確定度為Urel = 1%,k=2。
容量瓶:100 mL,不確定度為U=0.05 mL,k=2。
移液器:1000μL,相對不確定度為Urel = 0.3%,k=2。
光譜分析儀:不確定度為U = 0.5nm,k=2。
移液器分別取100μL、200μL、400μL水中尿素標準溶液,加入超純水定容至100 mL,標準溶液的質量濃度為1.0 mg/L、2.0 mg/L和4.0 mg/L。
3.1波長示值誤差
對一臺型號為UT-8900的尿素檢測儀進行測試,制造廠家為深圳市清時捷科技有限公司。
波長示值誤差、濃度示值誤差和濃度重復性的測試結果分別見表1、表2和表3。
本次測試對波長示值誤差和濃度示值誤差的測量不確定度進行評定,評定主要依據(jù)JJF 1822—2020《水樣檢測用尿素檢測儀校準規(guī)范》[2]和JJF 1059.1—2012《測量不確定度評定與表示》[3]。
5.1波長示值誤差測量不確定度的評定
5.1.1測量模型
5.3不確定度報告
對尿素檢測儀提出了三個測試項目,即波長示值誤差測試、濃度示值誤差的測試和濃度示值重復性,然后對樣機進行測試驗證和測量結果不確定度評定。該性能測試過程所使用的標準器和標準物質均可溯源,測試結果準確可靠。
【參考文獻】
[1]張倩倩.泳池水中尿素和氰尿酸現(xiàn)場快測方法研究[D].鄭州:鄭州大學,2022.
[2]水樣檢測用尿素檢測儀校準規(guī)范:JJF 1822—2020[S].
[3]測量不確定度評定與表示:JJF 1059.1—2012[S].
【作者簡介】
鐘妮,女,1986年出生,工程師,學士,研究方向為光學理化。
(編輯:李加鵬)