        Contralateral pneumothorax in the subacute phase after pacemaker implantation: lead retention and follow-up

        2019-01-16 09:07:22XiaoDongCHENXuXUJiaWenJIHaiLeiLIUJianPingSHEN
        Journal of Geriatric Cardiology 2018年12期

        Xiao-Dong CHEN, Xu XU, Jia-Wen JI, Hai-Lei LIU , Jian-Ping SHEN,

        1Department of Cardiology, Affiliated Hospital of Intergrated Traditional Chinese and Western Medicine, Nanjing University of Chinese Medicine,Nanjing, China

        2Jiangsu Province Academy of Traditional Chinese Medicine, Nanjing, China

        3Department of Radiology, Affiliated Hospital of Intergrated Traditional Chinese and Western Medicine, Nanjing University of Chinese Medicine,Nanjing, China

        4Department of Cardiology, the First Affiliated Hospital of Nanjing Medical University, Nanjing, China

        Keywords: Atrial lead perforation; Contralateral pneumothorax; Pacemaker implantation

        An 89-year-old woman [body mass index (BMI) = 20.27 kg/m2)] with hypertension underwent implantation of a dual-chamber pacemaker because of dizziness and syncope due to sick sinus syndrome. Pacemaker implantation via the left subclavian approach was performed on June 3, 2017. A right ventricular active lead (TendrilTM STS 2088TC, St Jude Medical, Penang, Malaysia) and an atrial active lead(OptisenseTM 1999, St Jude Medical, Penang, Malaysia)were placed in the right ventricular apex and anterior wall of the right atrium, respectively. Atrial lead parameters were normal (Table 1). The ventricular pacing threshold was 0.6 V at 0.5 ms, and the pacing impedance was 640 Ω. Sensed R wave was 17.2-18.4 mV. A postoperative chest radiograph showed normal findings with appropriate lead position (Figure 1A).

        Table 1. Atrial electrode parameters in Case 1.

        Figure 1. Images of chest X-ray and CT scan post pacemaker implantation during follow-up in case 1. (A): Five days after procedure with proper location of atrial and ventricular leads on X-ray; (B & C): extrusion of atrial lead helix through the right atrium (B, red arrow)caused pneumothorax and pneumopericaridum (C, red arrows) on CT scan nine days post procedure; and (D & E): no pneumothorax, pneu mopericaridum and pleural effusion still existed on CT scan at one-year follow-up.

        However, the patient was referred for assessment of nonspecific pain in the upper abdomen and mild dyspnea nine days after the operation. Auscultation revealed diminished respiration with dry and moist rales over the right chest, and a small amount of right-sided pneumothorax without obvious displacement of the electrodes was found on chest computed tomography (CT). Pneumopericardium,right pleural effusion, and extrusion of the screw-in atrial lead to the left lung, yet without gross displacement, were clearly observed on the chest CT scan (Figures 1B and 1C).Pacemaker interrogation results were as follows: slight changes in atrial lead parameters, a lower sensed P wave and decreased pacing impedance. Ventricular lead parameters remained stable. The patient’s symptoms relieved immediately after thoracic drainage. As the interrogation parameters were stable and chest radiography revealed resolution of pneumothorax in the following three days, the chest tube was removed subsequently and the atrial lead was not repositioned. During a 1-year follow-up period, the pacemaker parameters remained stable, and no residual pneumothorax, pneumopericardium, and pericardial effusion were found on CT scans (Figures 1D and 1E).

        In another case, a 93-year-old woman (BMI = 18.31 kg/m2) with hypertension underwent implantation of a dual-chamber permanent pacemaker for symptomatic sick sinus syndrome on June 26, 2017. Both atrial lead (Vitatron model, Crystalline Actfix ICQ09B-52B; Medtronic, Puerto Rico, USA) and ventricular lead (Vitatron model, Crystalline Actfix ICQ09B-58B, Medtronic) were inserted via the left subclavian vein and actively fixed in the right atrial appendage and right ventricular septum, respectively. Atrial pacing parameters were normal (Table 2). The ventricular pacing threshold was 0.6 V at 0.5 ms, with a pacing impedance of 640 Ω. Sensed R wave was 16.8-18.4 mV. Postoperative chest radiography showed normal findings with an appropriate lead position (Figure 2A).

        Seven days after the procedure, the patient experienced pain over the right chest and shortness of breath. Diminished respiration and moist rales were auscultated over the right chest. Chest radiography revealed moderate right-sided pneumothorax without obvious electrodes displacement.Chest CT was performed and revealed extrusion of the atrial lead into the left lung via the left atrial appendage (Figures 2B and 2C). Then, a chest tube was introduced, and the patient’s symptoms relieved promptly. As pneumothorax resolved and the atrial lead was stable with acceptable parameters, the chest tube was removed, and the atrial lead was retained. At the one-year follow-up, no discomfort wasreported. No residual pneumothorax and pericardial effusion were found on CT scanning (Figures 2D and 2E).

        Table 2. Atrial electrode parameters in Case 2.

        Figure 2. Images of X-ray and CT scan post pacemaker implantation during follow-up in case 2. (A): Five days after procedure with proper location of atrial and ventricular leads on X-ray; (B & C): extrusion of atrial lead helix through the right atrium caused pneumothorax(B, red arrow) and right pleural effusion (C, red arrow) on CT scan seven days post procedure; and (D & E): no pneumothorax and pleural effusion still existed on CT scan at one-year follow-up.

        Contralateral pneumothorax is a rare complication during pacemaker implantation,[1-6]and it mostly occurs in the acute phase (< 24 h) after pacemaker implantation because of active fixation leads perforating the lung via the right atrial wall. Subacute contralateral pneumothorax associated with atrial lead perforation is easily overlooked because of the phase during which it occurs,[1,5]and most patients undergo atrial lead repositioning.[7,8]We report two unusual cases of contralateral pneumothorax in elderly female patients with a low BMI in the subacute phase after pacemaker implantation, despite which the atrial leads were retained in situ.

        The risk factors for atrial perforation are as follows: old age (> 80 years), female sex, low BMI, an atrial lead positioned on the right free wall, corticosteroid therapy, lead repositioning, and so on.[1,2,5,9-11]Both our patients were elderly women with a low BMI who had atrial leads positioned in the right appendage or anterior wall of the right atrium. However, the reason for subacute contralateral pneumothorax despite stable atrial lead performance remains unknown. We speculated that because of the low profile of recent leads, the tip of the helix was micro-displaced, and perforated the pleura via the atrial wall, yet it plugged the small defect caused by perforation, avoiding pericardial effusion. Furthermore, cardiomyocyte contraction and fibrosis over the lead might play an important role in this phenomenon, preventing recurrence of pneumothorax after the subacute phase. However, in the early phase,the lead tip was not covered by fiber, and its repeated contact with the lung during inspiration gradually resulted in pneumothorax in the subacute phase. However, the electrode remained in contact with atrial myocardium, resulting in stable and normal lead parameters.

        Although no serious conditions were caused in our cases,we should try our best to avoid such complications as no long-term follow-up results have been reported and we surgical operations in such elderly patients might be risky.Firstly, thorough evaluation of the risk factors listed above is critical to guide pacemaker implantation. Secondly, the atrial septum should be preferred in high-risk patients. Alternatively, excessive pressure on the lead should be avoided when screwing in the atrial lead. During follow-up,chest radiography, echocardiography, pacemaker interrogation, and even chest CT should be performed in patients with right chest or upper abdomen pain and dyspnea. Thoracic drainage could help relieve symptoms if a large amount pneumothorax is confirmed. However, the atrial lead should be retained in the case of stable and acceptable atrial lead parameters. If the atrial lead parameters are unstable, the atrial lead should be repositioned through surgery.

        A micro-displaced atrial lead after pacemaker implantation could result in contralateral pneumothorax in the subacute phase, even without pericardial effusion or sudden changes in the pacemaker parameters. This condition can easily be identified by CT, and no progressive treatment is necessary besides chest intubation and close follow-up.

        Acknowledgment

        The authors declare that there are no conflicts of interest.

        av潮喷大喷水系列无码| 国产在线无码不卡影视影院| 国产农村妇女精品一二区| 激情另类小说区图片区视频区| 国产AV无码无遮挡毛片| 少妇下面好紧好多水真爽| 在线观看的网站| 伴郎粗大的内捧猛烈进出视频观看 | 人妻有码中文字幕| 成人精品一级毛片| 午夜日本理论片最新片| 亚洲精品国产熟女久久久| 夫妻免费无码v看片| 国产亚洲精品aaaaaaa片| 中文字幕久久久久久久系列| 水蜜桃在线观看一区二区国产| 欧美激欧美啪啪片| 欧美v亚洲v日韩v最新在线| 久久dvd| 国产的自拍av免费的在线观看| 真实国产精品vr专区| 国产乱码一二三区精品| 日日摸夜夜添夜夜添一区二区| 最近中文字幕一区二区三区| 在线观看在线观看一区二区三区| 国产尤物精品视频| 久久亚洲av永久无码精品| 国产91精品丝袜美腿在线| 久久久精品国产老熟女| 免费国产在线精品一区| 亚洲国产美女精品久久久久| 午夜一区二区三区在线视频| 日韩美女亚洲性一区二区| 极品av麻豆国产在线观看| 任你躁国产自任一区二区三区| 开心五月激动心情五月| 中文字幕在线观看| 少妇熟女视频一区二区三区| 风流少妇一区二区三区| 丰满女人猛烈进入视频免费网站| 中文字幕av一区中文字幕天堂 |