亚洲免费av电影一区二区三区,日韩爱爱视频,51精品视频一区二区三区,91视频爱爱,日韩欧美在线播放视频,中文字幕少妇AV,亚洲电影中文字幕,久久久久亚洲av成人网址,久久综合视频网站,国产在线不卡免费播放

        ?

        Quantitative Moisture Measurement with a Cavity Ring-down Spectrometer using Telecom Diode Lasers

        2015-01-13 02:25:32BingChenPengKngJinyingLiXioleiHeAnwenLiuShuimingHu
        CHINESE JOURNAL OF CHEMICAL PHYSICS 2015年1期

        Bing ChenPeng KngJin-ying LiXio-lei HeAn-wen LiuShui-ming Hu

        a.Hefei National Laboratory for Physical Sciences at the Microscale,University of Science and Technology of China,Hefei 230026,China

        b.Meteorological Observation Centre,China Meteorological Administration,Beijing 100081,China

        Quantitative Moisture Measurement with a Cavity Ring-down Spectrometer using Telecom Diode Lasers

        Bing Chena,Peng Kanga,Jian-ying Lib,Xiao-lei Heb,An-wen Liua,Shui-ming Hua?

        a.Hefei National Laboratory for Physical Sciences at the Microscale,University of Science and Technology of China,Hefei 230026,China

        b.Meteorological Observation Centre,China Meteorological Administration,Beijing 100081,China

        Moisture measurement is of great needs in semiconductor industry,combustion diagnosis, meteorology,and atmospheric studies.We present an optical hygrometer based on cavity ring-down spectroscopy(CRDS).By using diferent absorption lines of H2O in the 1.56 and 1.36μm regions,we are able to determine the relative concentration(mole fraction)of water vapor from a few percent down to the 10?12level.The quantitative accuracy is examined by comparing the CRDS hygrometer with a commercial chilled-mirror dew-point meter.The high sensitivity of the CRDS instrument allows a water detection limit of 8 pptv.

        Cavity ring down spectroscopy,Moisture,Trace detection,Diode laser

        I.INTRODUCTION

        Detection of moisture,or humidity in gases is of great needs in various felds.The concentration of water vapor should be below 1 ppmv(part per million by volume,mole fraction)in fabrication of semiconductor and photoelectric devices[1].There is a great demand of the capability detecting lower moisture levels in reactive gases used in semiconductor industry.Detecting trace moisture in fuels is also needed as a result of the increasing use of fuel ethanol.As a main product in combustion,concentration of water vapor is also a key factor in the diagnosis of fame and combustion processes.Determination of the humidity in ambient air is also of great importance in meteorology and atmospheric studies.

        Because of the ubiquity of water,its strong interactions with metals,and lacking of moisture-containing standard gases,it is difcult to quantitatively determine the water content in samples.Miscellaneous methods for moisture detection have been used in diferent applications.Dew-point hygrometers,which measure the water vapor condensation temperature[2],are widely used in meteorology and atmospheric studies[3]. Chilled-mirror hygrometers can detect a dew-point in the range from?100?C or so to the room temperature,which is equivalent to a water vapor pressure of 10?4?104Pa(Fig.1),with an accuracy of about 1%RH near the room temperature[4],but the precision gets much worse at very low water concentrations.An atmospheric pressure ionization mass spectroscopy could detect H2O in O2at the 1 ppbv(part per billion by volume)level[5].A capacitance-based sensor for detecting trace moisture in gases and organic solvents has also been recently reported[6].

        Optical methods based on the spectroscopy of water molecules attract more and more interests,especially in real-time trace moisture analysis[1].Fouriertransfer infrared(FTIR)spectroscopy is efective in multiple gases analysis and has been applied to detect moisture in electronic specialty gases[7].In the last two decades,laser-based spectroscopy techniques become more prominent.Tunable diode laser absorption spectroscopy(TDLAS)is relatively simple and robust, therefore has been widely applied[8-10].Cavity ringdown spectroscopy(CRDS)using cw diode lasers provides great enhancement in sensitivity and receives increasing interests in recent years[11-15].The reported detection limit of water vapor varies from a few ppbv

        FIG.1 Saturation vapor pressure of water calculated according to the formula given in Ref.[2].

        In this work,we present a CRDS instrument for quantitative trace moisture detection utilizing telecom distributed feedback(DFB)lasers.The absorption lines of H2O near 1.56 and 1.36μm are used to determine the water concentration in high and low humidity environments,respectively.Such an optical hygrometer can be used to measure water concentration from a few percent down to several pptv.The quantitative accuracy is verifed by comparing to a commercial chilled-mirror hygrometer.The instrument can achieve a water detection limit of 8 pptv,which is,to the best of our knowledge,improved by over one order of magnitude compared with those reported optical hygrometers.

        II.EXPERIMENTAL SETUP

        The confguration of the CRDS setup is similar to that presented in our previous work[16]and it is shown in Fig.2.A single-mode distributed feedback diode laser is used as the light source.The laser frequency is scanned in a step mode by controlling the injection current and the operation temperature.A beam of about 10 mW is coupled into a high-fnesse ringdown(RD)cavity.The cavity mirrors have a refectivity of 99.998%.The light emitted from the RD cavity is detected by an avalanche photo-detector.Once the detected signal is larger than a preset threshold,a trigger signal will be delivered to switch of the input laser beam using an acoustic-optical modulator(AOM) and the ring-down event will be recorded by a digitizer installed in a personal computer.On each step of the laser frequency,typically 20 ring-down events will be recorded.A ftting program is applied to ft the recorded data using an exponential function to obtain the decay timeτ.The absorption coefcient of the gas sample,α,can be derived from the equation,

        wherecis the speed of light,νis the laser frequency, andτandτ0are the decay time of the cavity with and without sample,respectively.

        An integration of detectedαaround a selected absorption line of water molecule should be proportional to the concentration of the molecules in the sample,

        whereSis the strength of the line of water in cm?1/(molecules·cm?2),andNis the number density of water molecules.Therefore,the percentage of the partial pressure of water in the gas sample can be determined according to the following equation,

        FIG.2 Experimental setup for humidity measurement with CRDS compared with a chilled-mirror dew-point meter. AOM:acoustic-optical modulator,OI:optical isolator.

        wherePtotalis the total pressure of the sample,NAis the Avogadro constant,Ris the molar gas constant, andTis the temperature of the gas sample.

        III.RESULTS

        A.Comparison with a dew-point meter,high humidity measurement

        The quantitative accuracy of the CRDS hygrometer was verifed by comparing with a chilled-mirror dewpoint meter.The experimental setup is presented in Fig.2.A humidity chamber(V¨otsch T/C 4-340,volume 335 L)is used to provide air samples with different concentrations of water vapor.The sensor of a chilled-mirror hygrometer(GE optica 1111H)is put in the chamber to continuously monitor the dew-point of the air sample.The ring-down cavity is connected to the humidity chamber using tefon tubes,and a dry pump is used to cycle the gas between the RD cavity and the humidity chamber.Note that both the temperature and pressure in the RD cavity are diferent from those in the humidity chamber,but the relative concentrations of water(χwater)should be the same. The total pressure in the humidity chamber is set to be 1 atm,while the pressure in the RD cavity is monitored by a capacitance manometer(MKS 627B)which has a relative accuracy of 0.15%.

        It is clearly shown in Fig.3 that the CRDS determined relative concentration of water has an excellent linear correlation to that given by the dew-point meter.However,there is a systematic deviation of about 4.4%between them.Since the dew-point meter has a relative accuracy of 1%and the statistical uncertainty of the CRDS measurement is below 0.5%,the systematic deviation between the CRDS and dew-point meter results most likely comes from the uncertainty in the strength of the water line.As stated in the HITRAN database[17],the relative uncertainty of the line strength is 5%?10%.If we use a corrected line strength of 2.738×10?25cm?1/(molecule·cm?2),instead of the value given in HITRAN,the relative deviation between the CRDS and dew-point meter results is only about 0.4%.

        When we changed the condition of the humidity chamber,we recorded the CRDS spectrum once every one minute,and a program was applied to ft the spectrum to calculate the integrated absorption coefcient in real time.Both the CRDS results and dew-point meter data have been recorded and are shown in Fig.4. Apparently,they agree with each other throughout the whole process,and it is also illustrated in Fig.4(b).The relative deviation between both methods is at a level below 1%,which agrees with the relative accuracy of the dew-point meter and the control precision of the humidity chamber.One can also notice that there is a little delay between the CRDS results and the dewpoint meter data.The reason is as the following∶when the condition is changing in the humidity chamber,the dew-point meter responses immediately since the sensor is located in the chamber,while the change will not be detected by CRDS until the sample in the ring-down cavity has been refreshed.The delay time,about one minute in our case,is related to the pumping rate and the volume of the RD cavity(about 1 L).

        FIG.3 Relative partial pressure of water vapor in the humidity chamber measured by CRDS and the dew-point meter.

        FIG.4(a)Five hours continuous measurement of the relative water concentration in a humidity chamber using a CRDS instrument and a dew-point meter.(b)Magnifed picture of the data obtained around 15.6 h.

        B.Low humidity,detection limit of water vapor

        In the case of very low water concentration beyond the capability of the humidity chamber,a different but stronger water line is selected for CRDS measurement to maintain a sufcient sensitivity and precision.Figure 5 shows a CRDS detected water line at 7331.2435 cm?1,which has a line strength of 1.238×10?22cm?1/(molecule·cm?2)given in the HITRAN database[17],about 450 times stronger than the line mentioned in Fig.3.The sample was pure nitrogen gas,continuously fowing through the RD cell. Using the line strength given in the HITRAN database, we can determine that the partial pressure of the water vapor in the cell was 0.395 Pa,corresponding to a water concentration of about 39.5 ppmv.

        The detection limit is proportional to the minimum detectable absorption coefcientαmin∶

        whereχminis the minimum detectable concentration of the water vapor,γis the proportional coefcient,Sis the strength of the selected absorption line of H2O,and?0is the height of the normalized line profle.In the case of a Gaussian profle,

        FIG.5 An absorption line of water at 7331.24 cm-1detected by the CRDS instrument.Water vapor presents as impurity in a nitrogen gas sample.The determined partial pressure of water is 0.40 Pa.Upper panel:observed and the simulated spectra,lower panel:ftting residuals.

        where?is the full width at halm maximum(FWHM). In the 1.36μm region,a typical strong water line has a strength of 1×10?20cm?1/(molecule·cm?2),and if a total gas pressure of about 10 kPa is applied,the line profle can be well described by a Gaussian function with a FWHM of about 0.02 cm?1.Therefore,we can get aγof about 0.8 cm.In this case,at a noise level ofαmin=1×10?10cm?1(as that shown in Fig.5),the corresponding water detection limit is 80 pptv.The sample pressure of 10 kPa is selected for optimized sensitivity and avoiding line overlapping.At a higher sample pressure,the pressure broadening could be dominant.If we cannot pump the RD cavity cell to the optimized pressure,for example,in anin situmeasurement,a broader spectrum should be recorded.If the selected water line does not overlap with other lines,the integrated line absorption can still be used for quantitative analysis.

        The detection limit of CRDS can be further improved by multiple spectrum averaging.Figure 6 shows an example of the averaged spectrum of H2O near 7335.7 cm?1.After averaging over 100 spectra recorded in about 2 h,the noise level decreases by one order of magnitude,from 1×10?10cm?1to 1×10?11cm?1.As a result,it corresponds to a water detection limit of 8 pptv.

        IV.CONCLUSION

        We built a cavity ring-down spectroscopy instrument based on telecom DFB fber lasers as an optical hygrometer.The concentration of water vapor is determined by measuring the absorption lines of H2O. Strengths of water lines in the 1.3?1.6μm region cover several orders of magnitudes,therefore we can select diferent water lines in the case of diferent humidities.A water line at 1.56μm,which has a line strength of 2.7×10?25cm/molecule,is used to measure the relative concentration of water vapor(mole fraction)of about 1%.By comparing to a commercial chilled-mirror dew-point meter,we have demonstrated that the CRDS results agree with the dew-point meter with a statistical deviation of 0.4%.Stronger water lines in the 1.36μm region are selected for measuring gases of lower humidity.A water line with a strength of 1.2×10?22cm/molecule is used to measure the mole fraction of water vapor at the ppmv level. The strongest water line in this region has a strength of 10?20cm/molecule,which allows us to extend the detection limit of water vapor to about 8 pptv at a sample pressure of 10 kPa.Such a detection limit corresponds to a partial pressure of H2O of 8×10?8Pa,or a molecular density of 2×107molecules/cm3.The optical hygrometer,which has a sensitivity comparable to mass spectrometers,but being more portable and less expensive,can be very useful in miscellaneous applications and studies.

        FIG.6(a)Averaged absorption spectrum of H2O near 7335.7 cm-1.(b)Minimum detectable absorption coefcient(αmin)from spectra with diferent averaging numbers (N).

        V.ACKNOWLEDGMENTS

        This work was supported by the National Basic Research Program of China(No.2013BAK12B00 andNo.2013CB834602)and the National Natural Science Foundation of China(No.21225314 and No.21427804).

        [1]H.H.Funke,B.L.Grissom,C.E.McGrew,and M.W. Raynor,Rev.Sci.Instrum.74,3909(2003).

        [2]B.Handy,The Proceedings of the Third International Symposium on Humidity and Moisture,London,England:Teddington,8(1998)

        [3]F.Hasebe,Y.Inai,M.Shiotani,M.Fujiwara,H. Voemel,N.Nishi,S.Y.Ogino,T.Shibata,S.Iwasaki, N.Komala,T.Peter,and S.J.Oltmans,Atmos.Chem. Phys.13,4393(2013).

        [4]N.D.Milosevic,N.M.Stepanic,and M.M.Babic, Therm.Sci.16,193(2012).

        [5]A.D.Scott,E.J.Hunter,and S.N.Ketkar,Anal. Chem.70,1802(1998).

        [6]S.I.Ohira,K.Goto,K.Toda,and P.K.Dasgupta, Anal.Chem.84,8891(2012).

        [7]H.H.Funke,M.W.Raynor,B.Yucelen,and V.H. Houlding,J.Electron.Mater.30,1438(2001).

        [8]S.Q.Wu,J.Morishita,H.Masusaki,and T. Kimishima,Anal.Chem.70,3315(1998).

        [9]C.S.Edwards,G.P.Barwood,P.Gill,B.Schirmer,H. Venzke,and A.Melling,Appl.Opt.38,4699(1999).

        [10]B.Schirmer,H.Venzke,A.Melling,C.S.Edwards,G. P.Barwood,P.Gill,M.Stevens,R.Benyon,and P. Mackrodt,Meas.Sci.Technol.11,382(2000).

        [11]S.Y.Lehman,K.A.Bertness,and J.T.Hodges,J. Cryst.Growth.250,262(2003).

        [12]V.Vorsa,S.Dheandhanoo,S.N.Ketkar,and J.T. Hodges,Appl.Opt.44,611(2005).

        [13]H.H.Funke,M.W.Raynor,K.A.Bertness,and Y. Chen,Appl.Spectrosc.61,419(2007).

        [14]P.A.Y.Fiadzomor,D.M.Baker,A.M.Keen,R.B. Grant,and A.J.Orr-Ewing,Appl.Spectrosc.62,1354 (2008).

        [15]H.Abe and K.M.T.Yamada,Sens.Actuat.A165, 230(2011).

        [16]B.Chen,Y.R.Sun,Z.Y.Zhou,J.Chen,A.W.Liu, and S.M.Hu,Appl.Opt.53,7716(2014).

        [17]L.S.Rothman,I.E.Gordon,Y.Babikov,A.Barbe, D.C.Benner,P.F.Bernath,M.Birk,L.Bizzocchi,V. Boudon,L.R.Brown,A.Campargue,K.Chance,E. A.Cohen,L.H.Coudert,V.M.Devi,B.J.Drouin, A.Fayt,J.M.Flaud,R.R.Gamache,J.J.Harrison, J.M.Hartmann,C.Hill,J.T.Hodges,D.Jacquemart,A.Jolly,J.Lamouroux,R.J.L.Roy,G.Li,D. A.Long,O.M.Lyulin,C.J.Mackie,S.T.Massie, S.Mikhailenko,H.S.P.Mller,O.V.Naumenko,A. V.Nikitin,J.Orphal,V.Perevalov,A.Perrin,E.R. Polovtseva,C.Richard,M.A.H.Smith,E.Starikova, K.Sung,S.Tashkun,J.Tennyson,G.C.Toon,V.G. Tyuterev,and G.Wagner,J.Quant.Spectrosc.Radiat. Transf.130,4(2013).

        10.1063/1674-0068/28/cjcp1410185to several hundred pptv(part per trillion by volume).

        ?Author to whom correspondence should be addressed.E-mail:smhu@ustc.edu.cn

        (Dated:Received on October 23,2014;Accepted on October 29,2014)

        尤物蜜桃视频一区二区三区| 成人午夜毛片| 亚洲av日韩片在线观看| 亚洲av自偷自拍亚洲一区| 91精品亚洲成人一区二区三区| 国内精品视频在线播放不卡| 精品露脸国产偷人在视频| 国模欢欢炮交啪啪150| 亚洲A∨无码国产精品久久网| av资源吧首页在线观看| 二区三区三区视频在线观看| 免费无遮挡禁18污污网站| 五月天久久国产你懂的| av中文码一区二区三区| 麻豆精品一区二区综合av| 性高湖久久久久久久久| 亚洲成年网站在线777| 精品国产夫妻自拍av| 五月开心婷婷六月综合| 国产97在线 | 中文| 国产午夜无码精品免费看动漫| 国产午夜视频高清在线观看 | 日本看片一区二区三区| 美利坚日韩av手机在线| 少妇无码太爽了不卡视频在线看| 国产偷国产偷高清精品| 一区二区三区日韩毛片| 亚州国产av一区二区三区伊在 | 亚洲熟妇乱子伦在线| 精品人妻午夜中文字幕av四季| 国产精品美女一区二区视频| 欧美与黑人午夜性猛交久久久| 亚洲乱码一区AV春药高潮| 天堂av网手机线上天堂| 成人美女黄网站色大免费的| 亚洲日本va99在线| 超碰青青草手机在线免费观看| 内射口爆少妇麻豆| 人妻在卧室被老板疯狂进入国产 | 成人片黄网站色大片免费观看cn | 国产日韩精品中文字幕|